Santa Barbara Infrared's new extended temperature blackbodies have a maximum absolute temperature of 175°C. The 2100 Series Blackbody System incorporates all the advantages of accuracy, stability, and ease of use as the 2000 Series Smart Blackbody®.
Main Features:
- Standard Temperature range: -15°C to +125°C delta T (10°C to 150°C Absolute)
- Optional Extended High Ranges: -20°C to +150°C delta T (5°C to 175°C Absolute)
- Setpoint resolution: . .001°C
- Display resolution: .001°C
Temperature Measurement Accuracy :
- Absolute Temperature: ±.001x(T-25°C) or .01°C, whichever is greater
- Stability: 0°C to 100°C, ±.004°C short term, ±.01°C long term175°C, ±.025°C short term, ±.04°C long term
- Emissivity : .985 ±.015, 2µ to 14µ
| |